•Biological safety cabin type: CLASS II according to EN 12469 standard.
•Work area and work surface built in AISI 304 stainless steel.
•Sliding front glass with alarm position. It can be raised or lowered manually or electronically, so that the cabin can be opened when it is turned off.
•The recycled air is filtered by an absolute HEPA H14 filter (99.995% MPPS EN1822 efficient) and is sent to the work area at a rate between 0.25 - 0.45 m/s
•Partial air recirculation according to regulations: 30% expulsion and 70% recirculation.
•The air expelled to the outside is filtered by an independent HEPA H14 absolute filter (99.995% MPPS efficiency EN1822 Standard) dedicated to this purpose.
•Work area maintains laminar flow conditions with an air purity type ISO 4 (according to ISO 14644) or Class 10 (according to Fed. Std. 209).
•Protective barrier in work area with a speed between 0.40 - 0.65 m/s.
•Microprocessor-controlled cabin with upgradeable FLASH memory and external display for easy visualization of cabin control parameters (speeds, flow rates, plenum pressure, expulsion barriers, humidity, temperature, etc...).
•The cab is equipped with an easy-to-use touch panel with safety key for white/ultraviolet lighting control, internal electrical socket.
•Alarms for detecting problems in the drive system and the protective barrier due to negative pressure and door opening, by voice and light warning on the screen.
•Fluorescent and UV exclusive lighting.
•Working voltage 110-220 V. 50/60 Hz. Maximum current consumption 4 A. (approx. 800 w.).
•Standard cab includes U.V. light; 2 Electrical outlets of 110-220 volts.